| 摘要:氯胺酮在临床上是一种高效麻醉剂,但近年来被滥用于各种娱乐场所,因此,建立一种快速、准确的氯胺酮检测方法的重要性日益凸显。对于现有的各种检测方法,都需要相应的样品前处理过程,增加了检测时间和检测复杂度,最后的检测结果一定程度上和样品前处理有着密切关系。本文根据现有技术的不足,使用超声雾化装置利用电子的高频振荡对样品进行雾化处理,将液体样品转化为水雾状,经过快速气相色谱分离,利用声表面波传感器的响应来检测物质。经过标准样品的检测,线性关系良好,相关系数R2在0.9481以上,检出限为0.1ug/ml,从样品处理到得到检测结果可在60s内完成。本系统样品前处理简单,可在一分钟内快速准确检测氯胺酮。通过标准溶液的实验,验证了本系统的可行性、可靠性和检测下限。 |
| 关键词: 快速气相色谱 氯胺酮检测 超声雾化 声表面波传感器 |
|
| On-site Detection of Ketamine Using the Ultrasonic Atomizer Combined with a Gas Chromatography-Surface Acoustic Wave Sensor |
|
|
| Abstract:As a high-efficient anesthetic, ketamine has been extensively abused for entertainment over the past few years. Therefore, to establish a method for quick and accurate determination of ketamine is certainly important. For most of the existing methods, pretreatment of ketamine samples is time-consuming and complex before determination. Meanwhile, the final results would be significantly influenced by the pretreatment as well, based on which this article introduces a novel method, using an ultrasound atomizer to convert liquid samples of ketamine into aerosol form and analyzing the sample with a self-built gas chromatography-surface acoustic wave sensor. Tested with the standard ketamine samples, this system showed a good linear correlation (R2=0.9481), with a determination limit of 0.1ug/ml. Besides, for a single sample, the determination time was reduced to 1 minute. This system was thus proved to be feasible, reliable and accurate for ketamine determination. |
| Key words: Fast GC Ketamine determination Ultrasonic atomization Surface acoustic wave sensor. |